|
You can open
the Feedback Details dialog box from the Senso Vita Feedback Test
window. The dialog box gives an overview of the feedback test. When you
have finished with the dialog box, select the
Close button to return to the
Feedback Test window.
The Feedback
Details dialog box displays the details about the feedback test. The
details for the measurement of the omnidirectional microphone setting
are displayed in the upper half of the dialog box, and the details for
the directional microphone setting, if available, are displayed in the
lower half of the dialog box.
Fb Test
The Fb Test
fields show whether the result of the feedback test was OK. If a field
shows asterisks, the test has failed in that frequency area.
Available Gain
The Available
Gain value is the measured feedback limit. Note that you cannot change
these parameters. They are measured values only.
Minimum
acceptable gain (MIN)
The minimum
acceptable gain value is the minimum gain required for normal speech for
a band. The available gain must be equal to or larger than the minimum
acceptable gain.
[TOP]
The Senso Vita
Feedback Test window is split into four panes, all relating to the
selected ear. Until you make a feedback test, no data are shown in the
window.
The two top
panes contain graphics for the feedback measurement. One of the panes
(right pane for the left ear) shows the Insertion Gain vs. Input curve
for the current fitting. The other pane shows an indication of the
actual feedback limits. Until you have made a feedback test, only the
MIN curve is shown. This curve indicates the minimum acceptable gain for
normal speech. When you have made a feedback test, bars indicating the
available gain values for the individual bands are shown.
The lower two
panes of the window show the parameters and the noise level indication.
In one of the panes you can see the noise level indication. As long as
the noise level in the room is not too high, the indication is green. If
the noise level rises above the acceptable limit, the noise indication
turns red.
In the last
pane you can see the measurement data for the feedback test. The Fb Test
parameter fields show dashes until you have made a feedback test. After
a test, the fields show whether the test is OK, that is, whether the
available gain is suitable for the frequency area in question. If a
field shows ***, the available gain is not satisfactory, and you should
consider remaking the earmould or shell.
If you need to
see the details of the feedback test, you can select the Details button
to open a window with more information.
[TOP]
The Feedback
Verification window is similar to the Feedback Test window for Senso
Vita. The upper two panes show the insertion gain vs. input graphics and
the feedback limit for the actual fitting.
In one of the
lower panes you can see the noise level indication, and in the last pane
you can see the minimum available gain (MIN) and the available gain
measured during the feedback test. Available gain data are displayed for
both microphone settings in the hearing aid.
-
Select
Feedback Verification from the
Tools menu. A window similar to the Feedback Test window opens
for the active ear. If a hearing aid is connected to the opposite ear,
this hearing aid is automatically muted as long as the Feedback
Verification window is open.
-
The noise bars indicate whether feedback
exists in the hearing aid. If a noise bar is red, feedback is present.
Select Apply to change the
available gain in the hearing aid in order to avoid feedback. You can
also change the Available Gain
parameter manually, like you change any other parameter in Compass.
-
You can set the hearing aid in a forced
max position by selecting one of the
Max radio buttons. There is a Max radio button for each of the
microphone settings.
-
If you need to make a new feedback test,
select the Start Test button.
This starts a feedback test for the active ear.
-
When you have finished the feedback
verification, select OK to
accept the changes and close the Feedback Verification window.
Note: You can use the button at
the bottom of the window to change to the Feedback Verification window
for the other ear. Select the red ear to change to the Right-ear
Feedback Verification window, and select the blue ear to change to the
Left-ear Feedback Verification window.
[TOP]
The Senso Diva Feedback Test window is
split into four panes all relating to the selected ear. Until you make a
feedback test, no data are shown in the window.
The two top panes contain graphics for
the feedback measurement. One of the panes (right pane for the left ear)
shows the Insertion Gain vs. Input curve for the current fitting. The
other pane shows an indication of the actual feedback limits. The bars
indicating the measured available gain values for the individual bands
are shown as soon as you have made a feedback test. In addition to the
bars, the MIN curve is shown. This curve indicates the minimum
acceptable gain for normal speech.
The lower two panes of the window show
the parameters and the noise level indication. In one of the panes you
can see the noise level indication. As long as the noise level in the
room is not too high, the indication is green. If the noise level rises
above the acceptable limit, the noise indication turns red.
In the last pane you can see the
measurement and target data for the feedback test. The four feedback
regions cover all bands in the Senso Diva hearing aids. The first row of
parameter fields (Fb Test) shows whether the test is OK, that is,
whether the available gain is suitable for the frequency area in
question. If a field shows *** the available gain is not satisfactory,
and you should consider remaking the earmould or shell.
The Minimum Acceptable Gain parameter
(MIN) covers the minimum gain requirement for normal speech, and the
Available Gain parameter values represent the result of the feedback
test.
-
Connect the hearing aids to the
interface.
-
Detect the hearing aids by
clicking Detect Hearing Aid or
selecting Detect Hearing Aid
from the Communication menu.
Compass detects the hearing aid type and serial numbers and displays
them on the screen.
-
Select
Fitting to accept the detected hearing aids and move to the
fitting procedure. Alternatively, you can select the desired models
manually.
-
A dialog box opens asking you which
hearing aids to activate and whether you want to use the Fitting
Wizard and the paediatric fitting rationale. If you choose to use the
Fitting Wizard, Compass takes you through the windows needed for the
fitting procedure. If you choose to make the fitting yourself, turn
off the Fitting Wizard by selecting the
No radio button.
-
Select
OK to open the Sensogram window for the right ear.
-
If you want to instruct your client
before you make a Sensogram, select the
Talk-over radio button. Before
you start making the Sensogram, activate the
Mute radio button again.
-
Measure Sensogram thresholds for the
four frequencies. Initially, the Sensogram HTL fields show dashes to
indicate that no measurement has been made. As soon as you select one
of the fields, the starting value is shown. A tone is presented to
your client when you select the
Activate Tone button. If you turn on the Auto Tone Activation
feature, a tone is presented whenever you change a parameter value.
-
When you have made a Sensogram for the
right ear, select Next to
accept the setting and open the Left-ear Sensogram window.
-
Measure a Sensogram for the left ear.
When finished, select Next to
close the window and open the Right-ear Feedback Test window.
-
Check that the noise level is not too
high, and select Start Test to
make a feedback test.
-
When the feedback test is finished, a
message box asks you to confirm it. Select
OK to close the message box.
-
Select
Next to open the Left-ear Feedback Test window.
-
Make a feedback test, and select
Next to move to the Program
Selection window.
-
Select the preferred program
combination, and select Finish
to complete the fitting.
-
You have now completed the basic
fitting. In the Complete Fitting dialog box, select
Fitting to move to the Fitting
window and see the measurements you have just made, or select
Fine Tuning if you want to go
directly to the Fine Tuning window.
[TOP]
When you open
the Fitting window for Senso Vita hearing aids, you can make a Sensogram
and a feedback test for the hearing aids.
The graphics in
the upper panes reflect the audiogram of the selected client. The darker
shade shown in the matrix marks the fitting range of the selected
hearing aid. If you have not measured a Sensogram, the Sensogram values
calculated by Compass are shown. If you have measured a Sensogram, the
measured values are shown in the graphics panes of the Fitting window.
Initially, the
parameter fields in the lower panes of the window only show the
Audiogram HTL values. Until you have made the necessary measurements,
the fields for the measurements only contain a dash.
You can shift
between the Mute and On radio buttons if you want to instruct your
client before you start making a Sensogram and/or a feedback test.
When you have
made the measurements, the four Sensogram fields show the measured
thresholds, and the three Fb Test fields show whether the feedback test
was OK in the three bands.
When you have
finished making measurements, you can move on to the Fine Tuning window
to work with the parameter values you obtain during the measurements.
[TOP]
The General
fine tuning window is the first window that opens when you enter fine
tuning for Senso Vita hearing aids.
The General
fine tuning window shows the Insertion Gain vs. Input graphics, but if
you need to see another type of graphics, you can change the view by
means of the Curve Types item on
the View menu.
In the General
window, you can change the IG Adjust parameters for all three bands in
the hearing aid. Use the arrow buttons next to the parameter fields to
change the setting. Note that whenever you change a parameter away from
the original setting, the changed parameter is marked by an exclamation
mark.
You can use the
Master Gain arrow buttons to the
right in the parameter panes to adjust the insertion gain for the normal
and loud input signals in all three bands simultaneously. Selecting the
up arrow increases all parameter values by 1 dB, and selecting the down
arrow decreases all parameter values by 1 dB.
From the
General fine tuning window you have access to windows focusing on the
different sound levels, and you have access to the Senso Vita Options
window. Use the horizontal navigation bar to open the other fine tuning
windows.
[TOP]
You can use the
Loud window to adjust the amplification of loud sounds, that is, for
Senso Vita you can adjust the Insertion Gain Adjust, Loud, parameter.
Use this window if your client has problems hearing loud sounds
correctly.
The window
shows the unaided and aided speech spectra for loud speech signals.
The lower part
of the window shows the IG Adjust, Loud, parameter, which you can
adjust. The Sensogram HTL values are also shown, but you cannot adjust
these values from the Loud window.
[TOP]
You can use the
Normal fine tuning window to adjust the amplification of normal speech,
that is, for Senso Vita you can adjust the Insertion Gain Adjust,
Normal, parameter. Use this window if your client has problems hearing
normal sounds correctly.
The window
shows the unaided and aided speech spectra for normal speech signals.
The lower part
of the window shows the IG Adjust, Normal, parameter, which you can
adjust. The Sensogram HTL values are also shown, but you cannot adjust
these values from the Normal window.
[TOP]
The features
available in the Options window depend on the type of hearing aid
selected. Below is a short description of the Senso Vita options. Click
a heading to get a more detailed description of the option.
PRG
The program
options let you select the desired combination of listening programs
used by the hearing aid. In the default setting the hearing aids use
Selection 1, but you can change the setting to match your client's
preferences.
Param
Under Program
Parameters you can define the settings of the Noise Manager, the Sound
Stabilizer, and the program gain in the hearing aid. Initially, the
program gain shows 0, but you can increase or decrease the setting.
DAI
Use the DAI tab
to see the setting of the Noise Manager and the Sound Stabilizer, and to
adjust the program gain for the direct audio input.
More
Under the More
tab you can define a few general features, such as the crossover
frequencies and the volume control.
[TOP]
During fine
tuning of Senso Diva and Senso Vita hearing aids, you have access to the
Program Checker window. This is a separate window, opening on top of the
active fine tuning window. The Program Checker window shows the symbols
for the programs available in the selected hearing
aid(s)
together with the features for which you can change the settings.
You can use the
program checker to activate the programs in the hearing aid. Use the
radio buttons below the symbols to activate the desired program. If you
need to mute the hearing aid while the program checker is active, you
can select the Mute radio button.
Using the
program checker
-
From any fine tuning window, select
Program Checker from the
Tools menu.
-
Select the radio button for the program
you want to work with.
-
Talk with your client, or play different
sounds, to find out whether the hearing aid program is set correctly.
-
If needed, change the settings of the
parameters available in the window.
-
When you have finished adjusting and
verifying the settings in the different programs, select the
Close button to close the
Program Checker window.
Compass returns
to the fine tuning window from which you opened the Program Checker and
sets the hearing aid in the master program.
The last step
in the basic fitting of a Senso Diva or a Senso Vita hearing aid is to
select the programs you want to use. The Program Selection window shows
the different program combinations you can use. The number and type of
programs available depend on the selected hearing aid model.
A hearing aid
can contain up to four programs. Below is a short description of each
program.
Quiet or M
Use the Quiet
or M (microphone) program for situations in which there is no or only
little surrounding noise.
Noise
Use the Senso
Vita Noise program for situations in which there is surrounding noise.
The hearing aid concentrates on sounds coming from the front and
eliminates sounds coming from the back.
Music
Use the Senso
Diva Music program for listening to music. In this program all settings
are optimised for listening to music.
M/T
Use the M/T
(microphone/telecoil)
program in situations in which both a telecoil and the microphone are
needed.
Telecoil
Use the
Telecoil program for situations in which only a telecoil is needed.
You can turn
this window off under Setup. If you use the Fitting Wizard and this
window is not displayed, the feature has been turned off.
The first
measurement you make during a fitting is a Sensogram.
The Sensogram
window is designed to approximate that of an audiometer. The Sensogram
HTL fields are comparable to the intensity dial or attenuator on an
audiometer, and the (Auto) Tone Activation field is similar to the tone
interrupter switch on an audiometer.
The Sensogram
window shows data in two panes relating to the same ear. The upper pane
shows a graphical view of the present settings, the lower pane shows the
parameters and, for Senso Diva, gives you access to the Expanded
Sensogram once you have measured thresholds in the four basic bands.
The graphical
part of the window shows the audiogram data (red or blue curve), the
fitting range, and the Sensogram graphs. Initially, the Sensogram graphs
show the values calculated by Compass, but when you measure the actual
thresholds, these are reflected in the graphics.
In the lower
part of the window you find the parameter fields for the Sensogram and
the Activate Tone button.
To make a
Sensogram, check that the correct hearing aids are selected, instruct
your client to respond to the tones presented, and do as follows:
-
Check that the hearing aids are placed
correctly on your client's ears, and check that the communication with
the interface works. Then open the Sensogram window.
-
Select a Sensogram HTL field to start
assessing a threshold for this band. You can change the value in the
Sensogram HTL field by using the arrow up and arrow down button next
to the parameter field, or by using the arrow keys on your keyboard.
Then select the Activate Tone
button to present the tone.
-
Present tones to your client to
establish a threshold. If Auto Tone Activation is turned on (there is
a check mark in front of the feature name), a signal is automatically
presented when you increase or decrease the tone intensity. When a
tone is activated, it is indicated on the Activate Tone button.
-
Repeat steps 2-3 for the other bands.
-
When you have finished the Sensogram for
one ear, select OK, or select
Next if you are running the
Fitting Wizard.
-
If you are making a binaural fitting,
repeat the Sensogram measurement for the opposite ear.
Note: When you have measured a
threshold in one band, simply leave the information on the screen and
repeat for the other bands.
Note: If you use the Auto Tone
Activation mode, Compass delays the presentation of a signal up to 2
seconds when switching between bands. This prevents your client from
reacting to your actions instead of reacting to the tones presented. If
you use this mode you must select an intensity in the Sensogram HTL
field when you change bands.
[TOP]
You can use the
Soft window to adjust the amplification of soft sounds, that is, for
Senso Vita you can adjust the Insertion Gain Adjust, Soft, parameter.
Use this window if your client has problems hearing soft sounds
correctly.
The window
shows the unaided and aided speech spectra for soft speech signals.
The lower part
of the window shows the IG Adjust, Soft, parameter, which you can
adjust. The Sensogram HTL values are also shown, but you cannot adjust
these values from the Soft window.
[TOP]
Widex
SoundTracker is a special view opening in a window on top of the main
Compass window. In this view, Compass continuously makes measurements
more than 15 times per second and displays the input and output on the
screen.
Note: You can only use
SoundTracker if you are working with a NOAHlink programming interface,
and if you are working online with Senso Diva or Senso Vita hearing
aids. You must also make a Sensogram and a feedback test for the hearing
aids before you can use SoundTracker.
Note:
The SoundTracker graphics are based on actual measurements in the
hearing aid. However, the displayed output is partly based on simulated
data, since the output signal is measured before the receiver. If you
enter individual RECD data into Compass, you will see even more accurate
performance graphics.
The view shows
the output versus frequency relationship. You can see the Sensogram
curve and the UCL curve as static curves. The input is shown in dB HL at
microphone level as the lightly coloured area in the view, and the
output is shown in dB HL at eardrum level as the deeper coloured area.
The maximum output within each second is also displayed as the coloured
'dotted line' above the output areas.
While
SoundTracker is open, you cannot change any parameters. Use the
Start SoundTracker button to
start measuring output. When SoundTracker is running you can use the
Stop SoundTracker button to stop
it and set the hearing aid on again.
You can also
choose whether you want to show the aided or the unaided output. Use the
radio buttons to choose the preferred view.
When you have
finished using SoundTracker, select the
Close button to close the window again.
[TOP]
In this view,
you can see how various speech spectra are processed through a Senso
Vita hearing aid. This view shows the following:
-
The
light shaded area represents the unaided speech spectrum or the
long-term average speech spectrum.
-
The
darker shaded area is the aided speech spectrum. This area
shows how various speech spectra are processed through the hearing
aid.
-
The
darker coloured markings reflect the Sensogram HTL values for
the right or left ear.
-
The
lighter coloured markings are the Sensogram UCL values for the
right or left ear.
-
Besides, the output curve is shown at
the relevant frequency. In the Normal window the output in dB HL at an
input of 65 dB SPL is shown, and in the Loud window the output in dB
HL at an input of 90 dB SPL is shown.
[TOP]
Senso Vita has
different test modes that you can use for different verification and
performance tests. When you activate a test mode it only remains active
as long as the hearing aid is connected to the interface and is online
in Compass.
For BTE models,
you can remove the programming adaptor, and the test mode setting
remains active in the hearing aid until you turn the hearing aid off and
on again.
Test Mode 1
This test mode
can be used to set Senso hearing aids in a special linear test mode that
should only be used for coupler measurements. The purpose of this test
mode is to allow reproduction of the Senso data for quality control,
service and so on.
The setting in
this test mode is totally independent of the fitting algorithm. The
hearing aid is set to linear mode in the current filter setting with a
gain level so that:
The hearing aid
is saturated when you make an OSPL90 measurement (input signal at 90 dB
SPL).
The hearing aid
is NOT saturated when you make a gain measurement with an input signal
at 50 dB SPL.
Warning:
This test mode should only be activated for coupler measurements. It
must NEVER be activated while the hearing aids are in your client's
ears.
Test Mode 2
This test mode
sets Senso in a special mode according to the fitting (it follows the
Insertion Gain vs. Input characteristics), but the noise detection
circuit and the statistically based regulation are de-activated.
[TOP]
|